In case of power failure (including the case that power
switch is turned off without stopping control) operations
after the recovery can be set.
Power recovery setting
1) Press
[
Set
]
key for longer than 5 seconds.
The mode enters into adjustment mode and
indicator displays [SnS] to enter into calibrating
displayed temperature mode.
2) Press
[▼]
key to enter into power recovery setting
mode
[
PoFF
]
.
3) Press
[
Set
]
key.
Indicator changes to display setting character
(blinking) of power recovery setting.
・「
Set key lamp
(
SV
)」
lights up.
※
Setting character (blinking) is the character that
was used previous time. Factory default is [cont].
4) Press
[▲]
or
[▼]
key to select setting character.
※
Setting character changes in order as below;
[
cont
]:
Continues to control
[▲]↓ ↑[▼]
[
oFF
]:
Stopping control
[▲]↓ ↑[▼]
[
on
]:
Continues to control. Timer recounts
from the setting value.
5) Press
[
Set
]
key. Setting character (blinking) is
determined and indicator displays [buSE] to enter
into buzzer setting mode.
※
Changed setting character will not be available until
you press
[
Set
]
key.
※
To change another adjustment mode, press [
▲
] or
[
▼
] key to change the mode.
※
If you do not change other adjustment mode, press
[Set] key for longer than 5 seconds. Indicator
changes to display measured temperature again.
※
If no key operation is done for longer than a minute
while setting, the display changes to measured
temperature display again. In such a case, please
reset all the settings from 1).
4
.
How to operate power recovery setting
1) Press
[
Set
]
key for longer
than 5 sec. to enter into adjustment mode.
2) Moves to power
recovery setting
3) Setting power
recovery mode
4) Setting power
recovery mode
5) Determining power recovery
setting. Moves to buzzer setting mode.
-44-
Press 3 times
PV
Run
Stop
Mode
Set
▼
▲
Def
h.m
SV
℃
Alarm
Heating
Auto stop
Auto start
1Step
Defrost
24Hour
Pro.
○
○
○
○
○
○
○
○
○
SnS
●
PV
Run
Stop
Mode
Set
▼
▲
Def
h.m
SV
℃
Alarm
Heating
Auto stop
Auto start
1Step
Defrost
24Hour
Pro.
○
○
○
○
○
○
○
○
○
PoFF
●
PV
Run
Stop
Mode
Set
▼
▲
Def
h.m
SV
℃
Alarm
Heating
Auto stop
Auto start
1Step
Defrost
24Hour
Pro.
○
○
○
○
○
○
○
○
cont
●
●
PV
Run
Stop
Mode
Set
▼
▲
Def
h.m
SV
℃
Alarm
Heating
Auto stop
Auto start
1Step
Defrost
24Hour
Pro.
○
○
○
○
○
○
○
○
oFF
●
●
PV
Run
Stop
Mode
Set
▼
▲
Def
h.m
SV
℃
Alarm
Heating
Auto stop
Auto start
1Step
Defrost
24Hour
Pro.
○
○
○
○
○
○
○
○
buSE
●
○