3-1. Control panel
㈰
㈮
㈯
㉀
㉂
㉃
㈺
㈱
㈹
㈪
㈫
㈬
㈭
㈷
-10-
h.m
SV
PV
Alarm
Heatin
g
Auto Stop
Auto Start
℃
Run
Stop
Mode
Set
▼
▲
3
Description of operating portion
Defrost
1Step
Def
Over Heat
Over Cool
Power switch
㈰
Turns on and off the instrument. When turning on this switch, indicator displays the
temperature of the chamber and fan in the air jacket starts working.
※
1
㈪
Mode key
Changes timer mode. Each time you press the key, the mode changes.
Pressing this key for longer than 5 sec. while timer working, it stops timer running.
※
2
Stationary operation
→
Auto stop
→
Auto star
t
→
1 step operation
→
24-hour operation
→ →
→
㈫
Defrost switch
㈬
Set key
Defrosts the condenser when pressing this switch.
Used for setting measured and setting value. Clears alarm while activating alarm.
(
except over heat and cool alarm
)
Stops buzzer when beeping.
㈭
Upper key
Available only for setting value or character. Each time you press the key, the value
increases by 0.1
℃(1
min.
)
,and holding down the key for longer than 1 second
can increase the value continuously.
※
3
When displaying the character, it changes each time you press the key.
㈮
Down key
Available only for setting value or character. Each time you press the key, the value
decreases by 0.1
℃(1
min.
)
,and holding down the key for longer than 1 second
can decrease the value continuously.
※
3
When displaying the character, it changes each time you press the key.
※
4
㈯
Run/Stop key
Starts and stops control.※5
㉀
Over heat alarm lamp
Lights up when over heat protector or over heat protection thermo stat works. Turns
off a light when the temperature in the chamber falls and continues to control
(recovers automatically).
㈷
Over cool alarm lamp
Lights up when over cool protector works. Turns off a light when the temperature in
the chamber falls and continues to control (recovers automatically).
㉂
Alarm lamp
Lights up when sensor or temperature control is in abnormal condition, or over load
on refrigeration unit, or power failure occurs, or door is open, or activating upper or
lower temperature alarm, or alarm for total operating hours occurs.
㉃
Heating lamp
Lights up when tuning on heater.
㈹
Refrigeration unit lamp
Lights up when turning on refrigeration unit.
㈺
Defrost lamp
Lights up when defrosting.