-12-
3-2
.
Stationary operation and timer mode
Timer mode is available in stationary operation
depending on the purpose of use. Input the setting on
control panel. Even half way of the operation, setting
temperature, time mode can be changed.
For more details about setting, please refer to
"Setting, running and stopping stationary operation"
on page 28 and "Setting, running and stopping timer
mode" on page 30.
Stationary operation
Time
Temperature
t
Time
Temperature
Time
Temperature
t
1. Stationary operation
If you set the temperature and start running the
unit, it performs continuous operation at the setting
temperature. Pressing
[
Run/Stop
]
key can start
and stop the operation.
※
Range of temperature setting
:
4.0
℃〜
60.0
℃
1. Auto stop mode
If you set the time
(t)
and start running the unit,
it performs stationary operation for setting time after
reaching at the setting temperature (setting value for
stationary operation). After elapsing the time, it
stops controlling automatically.
2.. Auto start mode
If you set the time (
t)
and start operation, the
unit performs stationary operation after elapsing
the time.
Timer mode
Constant number operation can be started and stopped
at the setting time of timer. Three types of timer modes are
available,
which are auto stop, auto start and 1-step operation.
Also, 24-hour cycle operation mode is available.
※
1. Auto stop, auto start, 1-step operation
Range of setting time for timer: 0
hr.
00
min.〜
99hrs.59min.
※
2. 24-hour cycle operation
Range of setting time for timer: 0hr.00min.
〜
23hrs.59min.
※
Each time you press the key, timer mode changes in
order as below;
Auto Stop
↓
Auto Start
↓
1 step operation
↓
24 Hour Pro.
↓
Temp.display (all the timer mode's LED turns off a
light).
Time
Temperature
3. 1-step operation mode
After reaching at the setting temperature (T1),
the unit performs stationary operation at preset
temperature (T2) after running for setting time
(t).
T1
T2
t