41
11.0 A pp en dix
11.0 Appendix
11.1 Technical data
The technical information and operating data can be found
in the current data sheet.
Data sheet for linear actuator for surgical tables and proce-
dure chairs – Series CAMT
PUB IL-07014-EN-October 2019
Current data sheets are available on the Ewellix website (see
ewellix.com).
Unit
CAMT20
Rated push load
N
6 000
Rated pull load
N
6 000
Static load (push/pull)
1)
N
13 200
Safety factor on rated load
2) 3)
–
4
Speed (full load to no load)
4)
mm/s
5 to 6,5
Stroke
mm
50 to 250
Voltage
VDC
24
Current consumption
A
10
Duty cycle
%
10 (1/9 minutes)
Ambient temperature
°C
+10 to +40
IP protection
–
IP20
Noise level (max)
dB
<= 55
Weight
5)
Kg
5,8
1)
Compliant with static load according to IEC/UL 60601-2-46
2)
Static safety factor to prevent mechanical hazards according to IEC/UL 60601-1
3)
Depending on stroke and attachment type, safe work load in push direction is reduced.
4)
Speed with 24 V DC, speed with V/SCU is higher.
5)
For stroke 250 mm, without attachment