CAMT 20
CAUTION
Damage due to wrong lubricants
The use of incorrect additives may cause significant material
damage
Therefore:
• Use only the auxiliary products listed by the manufacturer.
3. Insert DIN-8 plug (3) into the connecting socket (1) of the
control unit. Ensure correct positioning of the groove (2)
and the lug (4) (⮑ fig. 23).
Fig. 23
6.5.2 Connecting flying leads cable to
connector or external control unit
1. Follow the color coding of the leads defined in the tech-
nical details section (refer to chapter ⮑ 4.6. Connec-
tions, page 23).
2. Ensure that the control unit complies with the require-
ments for external control units defined in this operating
manual.
6.6 Connection to
operating element
NOTE
Operational conditions are not displayed on CAMT linear
actuators. However, this feature is mandatory in medical
applications and must be present on the control unit being used.
6.7 Connection to power
supply
NOTE
A control unit connects the device to the electric power supply
via a low voltage plug. Follow the control unit's operating manual
instructions during the connection process.
1. Connect linear actuator CAMT to control unit (refer to
chapter ⮑ 6.5 Connection to the control unit,
page 31).
2. Connect control unit to mains power supply (see related
operating manual for control unit).
3. Ensure that the plug of the power line is accessible at all
times.
6.8
Initial start-up
Before initial start-up, check that the following points have
been dealt with:
• All instructions followed in the above sections of this chapter.
• All cables secured against pinching and trapping, and
properly connected.
• Electric supply secured.
• Operating device connected to the control unit.
• No persons or obstacles near any movement.
• No mechanical hazard for all movements
Initial start-up
Prior to first operation, a professional electrician must per-
form and document the following test and readings:
• Visual condition check
• Function check of operating features and safety features
• Reading of leakage currents
• Reading of insulation resistance
After the installation check has been completed, it is possi-
ble to start up the device, the system.
Make sure that all functions are correctly operational.