DEPOLOX
®
400 M
Operation
6.
91
Please note
For a detailed description of the user administration, see “Menu “User ad
-
ministration”” on page93.
Max. dosing time
00:00 ... 10:00 h
This parameter defines a monitoring time which specifies the maximum time
the controllers may require before the measured value reaches the setpoint
minus monitoring hysteresis. If one of the controllers does not reach the set
-
point within the defined maximum dosing time, the corresponding controller
is switched off and a message is displayed. When the setting is “00:00:00,”
this function is switched off.
Please note: It is not possible in manual mode!
Monitoring hysteresis
2 ... 50 %
This parameter defines a range around the setpoint within which the set
-
point is recognized as reached when dosing time monitoring is enabled.
User administration
Access control
This parameter is used to switch the user administration on and off.
Off = no user administration; all parameters can be changed without enter
-
ing a password
On = user administration enabled, password protection or Level 1, 2 and 3
can be configured
Level 3
If access control is enabled, as least Level 3 must be used or configured.
Level 3 comprises read and write access to all setting parameters of the
device.
Unlock pattern
Menu for entry/definition of a Level 3 unlock pattern. The entry must be
repeated as confirmation.
Password
Menu for entry/definition of a Level 3 password. The entry must be
repeated as confirmation.
Level 2
If access control is enabled, Level 2 can be enabled or disabled. Level 2
allows access to setting parameters such as limit values, setpoint, date,
time and calibration. If Level 2 is enabled, an unlocking pattern and/or a
password must be defined.
Unlock pattern
Menu for entry of a Level 2 unlock pattern. The entry must be repeated as
confirmation.
Password
Menu for entry of a Level 2 password. The entry must be repeated as con
-
firmation.
Level 1
If access control is enabled, Level 1 can be enabled or disabled. Level 1
allows access to calibration. If Level 2 is not enabled, access to limit values
and setpoints is also possible on Level 1. If Level 1 is enabled, an unlock
-
ing pattern and/or a password must be defined.
Unlock pattern
Menu for entry/definition of a Level 1 unlock pattern. The entry must be
repeated as confirmation.
Password
Menu for entry/definition of a Level 1 password. The entry must be
repeated as confirmation.
Summary of Contents for WALLACE & TIERNAN DEPOLOX 400 M
Page 1: ...WALLACE TIERNAN DISINFECTION ANALYZER DEPOLOX 400 M Version 1 00 and later OPERATING MANUAL...
Page 2: ...2 WT 050 700 000 DE IM 0319 DEPOLOX 400 M Please note Original operating manual...
Page 36: ...36 WT 050 700 000 DE IM 0319 Description DEPOLOX 400 M 3...
Page 74: ...74 WT 050 700 000 DE IM 0319 Installation DEPOLOX 400 M 5...
Page 118: ...118 WT 050 700 000 DE IM 0319 Maintenance DEPOLOX 400 M 7...
Page 137: ...DEPOLOX 400 M Wiring diagrams 9 137 9 Wiring diagrams...
Page 138: ...138 WT 050 700 000 DE IM 0319 Wiring diagrams DEPOLOX 400 M 9...
Page 139: ...DEPOLOX 400 M Wiring diagrams 9 139...
Page 140: ...140 WT 050 700 000 DE IM 0319 Wiring diagrams DEPOLOX 400 M 9...
Page 142: ...142 WT 050 700 000 DE IM 0319 Explanations and Certificates DEPOLOX 400 M 10...
Page 143: ...DEPOLOX 400 M Explanations and Certificates 10 143 10 1 1 CSA certificate...
Page 144: ...144 WT 050 700 000 DE IM 0319 Explanations and Certificates DEPOLOX 400 M 10...
Page 145: ...DEPOLOX 400 M Explanations and Certificates 10 145...
Page 146: ...146 WT 050 700 000 DE IM 0319 Explanations and Certificates DEPOLOX 400 M 10...