60
WT.050.700.000.DE.IM.0319
Installation
DEPOLOX
®
400 M
5.
5.4.5
Installing the optional strainer
To install the optional strainer with ball valve (W3T158721) and pipe
clamp in the sample water supply line, proceed as follows:
1
Release the screw joint on the sample water inlet with ball valve (A)
(threaded connection G 1/2” A).
2
Connect strainer with pipe clamp (B).
3
Connect sample water inlet (C).
Image 4 Section, installation of strainer, straight
A Screw joint on sample water inlet with ball valve
B Strainer with pipe clamp
C Sample water inlet
5.4.6
Connecting the sample water outlet
Please note
No water pipes made of copper may be fitted in the installation. They
would falsify the measurement.
Sample water outlet in non-pressurized version
Proceed as follows:
1
On the non-pressurized version, no back-pressure is permitted in the
cell body.
2
The sample water outlet must be open. We recommend using a fun
-
nel above the outlet.
3
The sample water outlet must be routed in such a way that no siphon
effect can arise.
Sample water outlet in pressurized version
Proceed as follows:
1
On the pressurized version, a maximum back-pressure of 1.5 bar is
permitted on the sample water outlet.
2
Ensure that the drain screw (flow cell drain valve) is always closed.
Image 5 Section, flow cell module DEPOLOX
®
5 C, pressurized version
A Sample water inlet with ball valve
B Flow cell drain valve (drain at the drain screw)
C Sample water outlet with ball valve
A
B
C
A
B
C
Summary of Contents for WALLACE & TIERNAN DEPOLOX 400 M
Page 1: ...WALLACE TIERNAN DISINFECTION ANALYZER DEPOLOX 400 M Version 1 00 and later OPERATING MANUAL...
Page 2: ...2 WT 050 700 000 DE IM 0319 DEPOLOX 400 M Please note Original operating manual...
Page 36: ...36 WT 050 700 000 DE IM 0319 Description DEPOLOX 400 M 3...
Page 74: ...74 WT 050 700 000 DE IM 0319 Installation DEPOLOX 400 M 5...
Page 118: ...118 WT 050 700 000 DE IM 0319 Maintenance DEPOLOX 400 M 7...
Page 137: ...DEPOLOX 400 M Wiring diagrams 9 137 9 Wiring diagrams...
Page 138: ...138 WT 050 700 000 DE IM 0319 Wiring diagrams DEPOLOX 400 M 9...
Page 139: ...DEPOLOX 400 M Wiring diagrams 9 139...
Page 140: ...140 WT 050 700 000 DE IM 0319 Wiring diagrams DEPOLOX 400 M 9...
Page 142: ...142 WT 050 700 000 DE IM 0319 Explanations and Certificates DEPOLOX 400 M 10...
Page 143: ...DEPOLOX 400 M Explanations and Certificates 10 143 10 1 1 CSA certificate...
Page 144: ...144 WT 050 700 000 DE IM 0319 Explanations and Certificates DEPOLOX 400 M 10...
Page 145: ...DEPOLOX 400 M Explanations and Certificates 10 145...
Page 146: ...146 WT 050 700 000 DE IM 0319 Explanations and Certificates DEPOLOX 400 M 10...