26
WT.050.700.000.DE.IM.0319
Description
DEPOLOX
®
400 M
3.
Sample water outlet
On the pressurized version, the sample water supply is connected on the
output side via the G 1/2” connection on the ball valve. At this point, a
maximum back pressure of 1.5 bar is permitted.
On the non-pressurized version, the sample water runs off via a drain
connection for hoses with an internal diameter of 6 mm directly via the
cell body outlet at the top.
Flow cell drain valve
The flow cell drain valve can be used to collected a sample for calibration.
It is used to draw sample water from the cell body through the low-pres
-
sure side of the flow control valve or to drain the cell body for servicing
purposes.
Calibration holding clips
Two calibration holding clips are attached in the cover of the flow cell
module. For “hands-free” calibration of the sensors with buffer solution or
calibration solution (bag or beaker), the two calibration holding clips are
pushed into the side of the basic housing at the back.
Summary of Contents for WALLACE & TIERNAN DEPOLOX 400 M
Page 1: ...WALLACE TIERNAN DISINFECTION ANALYZER DEPOLOX 400 M Version 1 00 and later OPERATING MANUAL...
Page 2: ...2 WT 050 700 000 DE IM 0319 DEPOLOX 400 M Please note Original operating manual...
Page 36: ...36 WT 050 700 000 DE IM 0319 Description DEPOLOX 400 M 3...
Page 74: ...74 WT 050 700 000 DE IM 0319 Installation DEPOLOX 400 M 5...
Page 118: ...118 WT 050 700 000 DE IM 0319 Maintenance DEPOLOX 400 M 7...
Page 137: ...DEPOLOX 400 M Wiring diagrams 9 137 9 Wiring diagrams...
Page 138: ...138 WT 050 700 000 DE IM 0319 Wiring diagrams DEPOLOX 400 M 9...
Page 139: ...DEPOLOX 400 M Wiring diagrams 9 139...
Page 140: ...140 WT 050 700 000 DE IM 0319 Wiring diagrams DEPOLOX 400 M 9...
Page 142: ...142 WT 050 700 000 DE IM 0319 Explanations and Certificates DEPOLOX 400 M 10...
Page 143: ...DEPOLOX 400 M Explanations and Certificates 10 143 10 1 1 CSA certificate...
Page 144: ...144 WT 050 700 000 DE IM 0319 Explanations and Certificates DEPOLOX 400 M 10...
Page 145: ...DEPOLOX 400 M Explanations and Certificates 10 145...
Page 146: ...146 WT 050 700 000 DE IM 0319 Explanations and Certificates DEPOLOX 400 M 10...