58
WT.050.700.000.DE.IM.0319
Installation
DEPOLOX
®
400 M
5.
5.4.3
Removing and fitting the housing cover
1
Remove the housing cover of the flow cell module. To do this, press
both unlocking buttons on the top of the housing and carefully
remove the cover toward the front.
2
Fit and engage the housing cover of the flow cell module. To do this,
position the housing cover at the bottom and carefully press it
upward until the housing cover engages on the unlocking buttons.
Image 1 View of top of housing cover
A Unlocking buttons
5.4.4
Connecting the sample water inlet
Please note
No water pipes made of copper may be fitted in the installation. They
would falsify the measurement.
When connecting the sample water inlet, note the following:
• A distinction must be made between a sample water inlet with tubing
connection and a sample water inlet with rigid pipework.
• The pressure in the sample water inlet must always be within a range
of min. 0.25 to max. 3.0 bar. The pressure in the sample water inlet
must always be 0.25 bar higher than in the sample water outlet.
• If the admission pressure is below 0.25 bar, a booster pump must be
used.
• If the admission pressure exceeds 3.0 bar, an optional pressure
reducing valve must be used.
• To prevent long loop lag times, ensure that the lines in the sample
water inlet are as short as possible.
• An optional strainer (W3T158721) with a mesh size of 0.5 mm can be
provided for the sample water inlet.
Image 2 Section, flow cell module DEPOLOX
®
5 C, pressurized version
A Sample water inlet with ball valve
B Flow cell drain valve (drain at the drain screw)
C Sample water outlet with ball valve
A
A
A
B
C
Summary of Contents for WALLACE & TIERNAN DEPOLOX 400 M
Page 1: ...WALLACE TIERNAN DISINFECTION ANALYZER DEPOLOX 400 M Version 1 00 and later OPERATING MANUAL...
Page 2: ...2 WT 050 700 000 DE IM 0319 DEPOLOX 400 M Please note Original operating manual...
Page 36: ...36 WT 050 700 000 DE IM 0319 Description DEPOLOX 400 M 3...
Page 74: ...74 WT 050 700 000 DE IM 0319 Installation DEPOLOX 400 M 5...
Page 118: ...118 WT 050 700 000 DE IM 0319 Maintenance DEPOLOX 400 M 7...
Page 137: ...DEPOLOX 400 M Wiring diagrams 9 137 9 Wiring diagrams...
Page 138: ...138 WT 050 700 000 DE IM 0319 Wiring diagrams DEPOLOX 400 M 9...
Page 139: ...DEPOLOX 400 M Wiring diagrams 9 139...
Page 140: ...140 WT 050 700 000 DE IM 0319 Wiring diagrams DEPOLOX 400 M 9...
Page 142: ...142 WT 050 700 000 DE IM 0319 Explanations and Certificates DEPOLOX 400 M 10...
Page 143: ...DEPOLOX 400 M Explanations and Certificates 10 143 10 1 1 CSA certificate...
Page 144: ...144 WT 050 700 000 DE IM 0319 Explanations and Certificates DEPOLOX 400 M 10...
Page 145: ...DEPOLOX 400 M Explanations and Certificates 10 145...
Page 146: ...146 WT 050 700 000 DE IM 0319 Explanations and Certificates DEPOLOX 400 M 10...