Operation
6.
81
pH measurement
The following parameter settings are possible for pH measurement:
Measurement
Range
Range end
pH 0.00 to pH 14.00
Range start
pH 0.00 to pH 14.00
Measurement filter
off/low/middle/strong
Limits
Limit values I
Max
Within range
Min
Within range
Hysteresis
1 to 25
Limit values II
Max
Within range
Min
Within range
Hysteresis
1 to 25
Maintenance intervals
Calibration
Message
On/Off
Interval
2 to 90
next maintenance on
Date
Reset maintenance
Execute
Cell
Message
On/Off
Interval
10 to 365
next maintenance on
Date
Reset maintenance
Execute
Summary of Contents for WALLACE & TIERNAN DEPOLOX 400 M
Page 1: ...WALLACE TIERNAN DISINFECTION ANALYZER DEPOLOX 400 M Version 1 00 and later OPERATING MANUAL...
Page 2: ...2 WT 050 700 000 DE IM 0319 DEPOLOX 400 M Please note Original operating manual...
Page 36: ...36 WT 050 700 000 DE IM 0319 Description DEPOLOX 400 M 3...
Page 74: ...74 WT 050 700 000 DE IM 0319 Installation DEPOLOX 400 M 5...
Page 118: ...118 WT 050 700 000 DE IM 0319 Maintenance DEPOLOX 400 M 7...
Page 137: ...DEPOLOX 400 M Wiring diagrams 9 137 9 Wiring diagrams...
Page 138: ...138 WT 050 700 000 DE IM 0319 Wiring diagrams DEPOLOX 400 M 9...
Page 139: ...DEPOLOX 400 M Wiring diagrams 9 139...
Page 140: ...140 WT 050 700 000 DE IM 0319 Wiring diagrams DEPOLOX 400 M 9...
Page 142: ...142 WT 050 700 000 DE IM 0319 Explanations and Certificates DEPOLOX 400 M 10...
Page 143: ...DEPOLOX 400 M Explanations and Certificates 10 143 10 1 1 CSA certificate...
Page 144: ...144 WT 050 700 000 DE IM 0319 Explanations and Certificates DEPOLOX 400 M 10...
Page 145: ...DEPOLOX 400 M Explanations and Certificates 10 145...
Page 146: ...146 WT 050 700 000 DE IM 0319 Explanations and Certificates DEPOLOX 400 M 10...