12
WT.050.700.000.DE.IM.0319
Description
DEPOLOX
®
400 M
3.
3.2
Versions
The following modules are available:
Overview of sensors
The table below contains an overview of the sensors. For further details
of the sensors, see Chapter 3.9 “Technical data”.
Part No.
Name
Electronic module 400 M (Module name E02)
W3T441932
Electronic module 400 M with 2-channel disinfection
analyzer and 4-way mA output module (factory
installed)
Flow cell module (Module name D02)
W3T389205
Flow cell module DEPOLOX
®
5 C, non-pressurized version
W3T389210
Flow cell module DEPOLOX
®
5 C, pressurized version
W3T440600
Flow cell module VariaSens C, non-pressurized version
W3T440611
Flow cell module VariaSens C, pressurized version
Y-style flow-through assembly
W3T167442
Y-style flow-through assembly, non-pressurized version
Sensor measuring modules
W3T441935
Sensor measuring module TC3 (free chlorine)
W3T441936
Sensor measuring module FC2 (total chlorine)
W3T443072
Sensor measuring module CD7 (chlorine dioxide)
W3T443073
Sensor measuring module OZ7 (ozone)
W3T441938
Sensor measuring module pH
W3T441937
Sensor measuring module fluoride
Name
3-electrode cell
(DEPOLOX
®
5 C)
TC3 (total
chlorine)
FC2
(free chlo
-
rine)
CD7
(chlorine
dioxide)
OZ7
(ozone)
Measuring
principle
Potentiostatic 3-elec
-
trode cell with plati
-
num electrodes
Membrane-covered, ampero
-
metrically operating potentio
-
static 3-electrode system with
integrated electronics
Membrane-covered
2-electrode system
Measured
value
Free Cl
2
, ClO
2
, O
3
,
KMnO
4
, Cl
2
++ for pH-
compensated free Cl
2
measurement
Total chlorine
Free chlorine Chlorine
dioxide (ClO
2
)
Ozone (O
3
)
Area of
application
Drinking water, high
level of hardness and
fast reaction time
Drinking water, suitable for low
conductivity and fluctuating pH
value
In water with dirt particles <
0.8 mm
Summary of Contents for WALLACE & TIERNAN DEPOLOX 400 M
Page 1: ...WALLACE TIERNAN DISINFECTION ANALYZER DEPOLOX 400 M Version 1 00 and later OPERATING MANUAL...
Page 2: ...2 WT 050 700 000 DE IM 0319 DEPOLOX 400 M Please note Original operating manual...
Page 36: ...36 WT 050 700 000 DE IM 0319 Description DEPOLOX 400 M 3...
Page 74: ...74 WT 050 700 000 DE IM 0319 Installation DEPOLOX 400 M 5...
Page 118: ...118 WT 050 700 000 DE IM 0319 Maintenance DEPOLOX 400 M 7...
Page 137: ...DEPOLOX 400 M Wiring diagrams 9 137 9 Wiring diagrams...
Page 138: ...138 WT 050 700 000 DE IM 0319 Wiring diagrams DEPOLOX 400 M 9...
Page 139: ...DEPOLOX 400 M Wiring diagrams 9 139...
Page 140: ...140 WT 050 700 000 DE IM 0319 Wiring diagrams DEPOLOX 400 M 9...
Page 142: ...142 WT 050 700 000 DE IM 0319 Explanations and Certificates DEPOLOX 400 M 10...
Page 143: ...DEPOLOX 400 M Explanations and Certificates 10 143 10 1 1 CSA certificate...
Page 144: ...144 WT 050 700 000 DE IM 0319 Explanations and Certificates DEPOLOX 400 M 10...
Page 145: ...DEPOLOX 400 M Explanations and Certificates 10 145...
Page 146: ...146 WT 050 700 000 DE IM 0319 Explanations and Certificates DEPOLOX 400 M 10...