WT.050.560.001.UA.IM.0714
36
Depolox
®
3 plus ReSIDUAl ANAlYZeR
eVoQUA
W3T140188
3.1.3.2
Calibration Menu
Display
Value Range
(defaults bold)
Description
CALIBRATION
Calibration-Menu
All calibration settings take place via this menu path.
DIS zero
range
Disinfection - zero adjustment
Pressing the Check-button sets the disinfection reading to
zero.
DIS span
range
Disinfection - span adjustment
You put in the result of the Comparison measurement to cali-
brate the disinfection span.
offset pH
-1.00 pH
+1.00 pH
0.00 pH
pH-Offset-Adjustment
(only in case of pH-selection)
This value is added to the actual pH reading.
calibr. at pH7
range
pH-Calibration
(only in case of pH-selection)
First point of the pH-calibration.
calibration pH
range
pH-Calibration
(only in case of pH-selection)
Second point of the pH-calibration.
offset fluor
-1.00 mg/l
+1.00 mg/l
0.00
mg/l
Fluoride-Offset-Adjustment
(only in case of fluoride-selection)
This value is added to the actual fluoride reading.
Fluor zero
range
Fluoride-Calibration
(only in case of fluoride-selection)
First point of the fluoride-calibration (two point
calibration).
Fluor span
range
Fluoride-Calibration
(only in case of fluoride-selection)
Second point of the fluoride-calibration (two point
calibration).
calibration temp
20.0°C
-5.0°C
+5.0°C
Calibration of sample water temperature
You can adjust the reading of the sample water
measurement (-5...+5°C).
Summary of Contents for Wallace & Tiernan Depolox 3 plus
Page 11: ...WT 050 560 001 UA IM 0714 Depolox 3 plus RESIDUAL ANALYZER EVOQUA W3T140188 ...
Page 12: ......
Page 19: ......
Page 39: ...WT 050 560 001 UA IM 0714 26 Depolox 3 plus RESIDUAL ANALYZER EVOQUA W3T140188 ...
Page 40: ......
Page 109: ......
Page 127: ...WT 050 560 001 UA IM 0714 112 Depolox 3 plus RESIDUAL ANALYZER EVOQUA W3T140188 ...
Page 128: ......
Page 146: ...WT 050 560 001 UA IM 0714 130 Depolox 3 plus RESIDUAL ANALYZER EVOQUA W3T140188 ...
Page 147: ......
Page 157: ......