24. Screw the cap back on.
25. Fill half a cap from the plastic bottle with cell sanc and pour it into the
cell body (approx. 1/2 cm³ cell sand) (see section 3.3.4 “Adding Grit
(Bare Electrode)”).
26. Reinsert electrodes.
27. Reopen the check valve on the sample water inlet.
28. Perform calibration after approximately three hours running-in time.
NOTE: Perform a calibration after one hour running-in time and, if required,
after 24 hours. you must calibrate each time the cell sand is replaced.
The calibration must be checked after one day. Note the electrolyte’s
expiration date. We recommend replacing the cell sand when replacing
the electrodes and porous membranes.
Cleaning/Replacing the Fine Filter
A fine filter must only be installed when membrane sensors are employed.
The fine filter must be cleaned or replaced in order to protect the mem-
brane sensor’s delicate membrane against soiling or damage.
1. Release both knurled nuts (A)
2. Remove complete filter unit (D)
3. Remove the fine filter (C)
A
Knurled nut
B
O-ring
C
Fine filter
D
Filter unit
Summary of Contents for Wallace & Tiernan Depolox 3 plus
Page 11: ...WT 050 560 001 UA IM 0714 Depolox 3 plus RESIDUAL ANALYZER EVOQUA W3T140188 ...
Page 12: ......
Page 19: ......
Page 39: ...WT 050 560 001 UA IM 0714 26 Depolox 3 plus RESIDUAL ANALYZER EVOQUA W3T140188 ...
Page 40: ......
Page 109: ......
Page 127: ...WT 050 560 001 UA IM 0714 112 Depolox 3 plus RESIDUAL ANALYZER EVOQUA W3T140188 ...
Page 128: ......
Page 146: ...WT 050 560 001 UA IM 0714 130 Depolox 3 plus RESIDUAL ANALYZER EVOQUA W3T140188 ...
Page 147: ......
Page 157: ......