WT.050.560.001.UA.IM.0714
96
Depolox
®
3 plus ReSIDUAl ANAlYZeR
eVoQUA
W3T140188
4.1
Depolox® 3
plus
Table 4.1 - errors
Error Message
Cause
Remedy
High/Low Alarms:
*min DIS ? *
Alarm value exceeded
Check dosing
*max DIS ? *
Check sample water flow
*min pH ? *
*max pH ? *
*min Fluor ? *
*max Fluor ? *
General Faults:
*mA-output? *
mA loop impedance too high
Check mA loop <1000 ohm
Loop interrupted
Jump if not used
*cell DIS ? *
Sensor wrongly connected
Check wiring
*cell pH ? *
Sensor cables interchanged
or
(In ph compensated mode) pH is
out of range
Perform calibration
or
Adjust pH
*cell Fluor ? *
Sensor defective
Replace sensor
*Circ.Fail. ? *
Internal failure
Contact Evoqua Water
Technologies
*temperature ? *
Temperature failure
Check temp. Sensor, wiring
Warnings
*Range ? *
Alarm value out of range
Adjust range or limit value
Range changed subsequently
Setpoint Cl
2
out of range
Adjust range or setpoint
*ADU 1 ? *
Internal failure
Call Evoqua Water
Technologies service
*ADU 2 ? *
*ADU 3 ? *
*Cal. DIS ? *
Calibration error
Perform new calibration
*Cal. pH ? *
Check buffer solutions
*Cal. Fluor ? *
Replace electrolyte
*OVR DIS ? *
Value exceeds range
Check range and dosing
*OVR pH ? *
Range does not fit
*OVR Fluor ? *
Dosing too high
*DI1 ? *
*DI2 ? *
*DI3 ? *
Signal at the digital input
Check for the origin of the signal,
e.g., sample water flow too low
Jump if not used
Summary of Contents for Wallace & Tiernan Depolox 3 plus
Page 11: ...WT 050 560 001 UA IM 0714 Depolox 3 plus RESIDUAL ANALYZER EVOQUA W3T140188 ...
Page 12: ......
Page 19: ......
Page 39: ...WT 050 560 001 UA IM 0714 26 Depolox 3 plus RESIDUAL ANALYZER EVOQUA W3T140188 ...
Page 40: ......
Page 109: ......
Page 127: ...WT 050 560 001 UA IM 0714 112 Depolox 3 plus RESIDUAL ANALYZER EVOQUA W3T140188 ...
Page 128: ......
Page 146: ...WT 050 560 001 UA IM 0714 130 Depolox 3 plus RESIDUAL ANALYZER EVOQUA W3T140188 ...
Page 147: ......
Page 157: ......