WT.050.560.001.UA.IM.0714
16
Depolox
®
3 plus ReSIDUAl ANAlYZeR
eVoQUA
W3T140188
2.6
Flow Switch (Optional) (Depolox
®
3
plus
Wet Side Only)
In general, the flow switch is only connected to the digital input of the Depolox®
3
plus
. In case of other applications, ensure that the rating of this Reed contact
(100 V AC and 0.5 A) is not exceeded even for a short time. In case inductive
loads (e.g., relays or contactors) are switched, the contact has to be protected
against induction sparks.
Figure 2.3 - Flow Switch
See Dwgs. 50.560.100.020 and .030 for flow switch location.
2.7
Stop Valve (Optional) (Depolox® 3
plus
Wet Side Only)
Figure 2.4 - Stop Valve
See Dwgs. 50.560.100.020 and .030 for stop valve location.