WT.050.560.001.UA.IM.0714
103
Depolox
®
3 plus ReSIDUAl ANAlYZeR
eVoQUA
W3T140188
(1) Counter electrode (blue spot) blue wire
(4) earth grey wire
(2) Working electrode (red spot) pink wire
(5) Temperature sensor brown wire
(3) Reference electrode white wire
(6) Temperature sensor yellow & green wire
Figure 4.1
m. Reassemble the cover (10) and cable gland (11) to the electrode hous-
ing.
n. Refit the electrode/electrolyte reservoir assembly and secure with the
clamping nut (2).
o. Reconnect the water inlet line, outlet line and the cell drain line.
p. Leave the cell running for an hour, to allow the system to settle down,
and then carry out the zero calibration procedure. This procedure may
be repeated after 24 hours if required.
4
5
6
3
1
2
Summary of Contents for Wallace & Tiernan Depolox 3 plus
Page 11: ...WT 050 560 001 UA IM 0714 Depolox 3 plus RESIDUAL ANALYZER EVOQUA W3T140188 ...
Page 12: ......
Page 19: ......
Page 39: ...WT 050 560 001 UA IM 0714 26 Depolox 3 plus RESIDUAL ANALYZER EVOQUA W3T140188 ...
Page 40: ......
Page 109: ......
Page 127: ...WT 050 560 001 UA IM 0714 112 Depolox 3 plus RESIDUAL ANALYZER EVOQUA W3T140188 ...
Page 128: ......
Page 146: ...WT 050 560 001 UA IM 0714 130 Depolox 3 plus RESIDUAL ANALYZER EVOQUA W3T140188 ...
Page 147: ......
Page 157: ......