WT.050.560.001.UA.IM.0714
48
Depolox
®
3 plus ReSIDUAl ANAlYZeR
eVoQUA
W3T140188
ror condition continues, it may be necessary to adjust the cell volt-
age Upot: Go to SIGNAL DIAG. and menu “Cl
2
cell current”.
NOTE: New units are factory calibrated. No Upot adjustments are
required on new units during initial calibrations.
If the display still does not read 0.00 mg/l after calibration, adjust
the potentiometer Upot so that the displayed cell current is ±0.5 µA
(turning clockwise reduces the current) (see Figure 3.5). The current
will change very slowly. Wait one half hour and check again. Perform
calibration again.
Figure 3.5 - Upot Adjustment
e. Let the sample water flow again.
NOTE: When there is no key pressed for five minutes, the Depolox® 3
plus
returns to the basic display.
• Span calibration
a. After zero calibration wait at least two minutes. Then take a sample at
the measuring cell. Determine the residual value of this sample.
b. Press the
key until the menu “DIS span” is reached. Press the
key
to open the menu.
c. Press keys
or
until the displayed value agrees with the manually
measured value.
d. Store the value using the
key. The chlorine measuring cell is now
calibrated. In case of an error message, refer to paragraph 4.1.1, Errors.
Summary of Contents for Wallace & Tiernan Depolox 3 plus
Page 11: ...WT 050 560 001 UA IM 0714 Depolox 3 plus RESIDUAL ANALYZER EVOQUA W3T140188 ...
Page 12: ......
Page 19: ......
Page 39: ...WT 050 560 001 UA IM 0714 26 Depolox 3 plus RESIDUAL ANALYZER EVOQUA W3T140188 ...
Page 40: ......
Page 109: ......
Page 127: ...WT 050 560 001 UA IM 0714 112 Depolox 3 plus RESIDUAL ANALYZER EVOQUA W3T140188 ...
Page 128: ......
Page 146: ...WT 050 560 001 UA IM 0714 130 Depolox 3 plus RESIDUAL ANALYZER EVOQUA W3T140188 ...
Page 147: ......
Page 157: ......