WT.050.560.001.UA.IM.0714
84
Depolox
®
3 plus ReSIDUAl ANAlYZeR
eVoQUA
W3T140188
3.7.4.2
Request - Frame
This is necessary to read data or additional information from a slave.
Format of the request frame:
Byte: 0-2
3
4
5
6
7
8
9
——————————————————————
Name: SYN
SB
SA
ZA
KB
AB
FC
EB
Byte:
Name:
occupied by:
0-2
SYN
Synchronization bytes
3
SB
Start byte 10H
4
SA
Slave address
5
ZA
destination address
6
KB
check byte
7
AB
number of bytes
8
FC
frame Check
9
EB
end byte 16H
With the request frame, single words, values covering more addresses, or ad-
ditional information of the destination address can be read. If the contents of
a single destination address are requested, the number byte is set to 00Hex. In
the answer frame, the data format and the number of bytes are entered from
the address reference list to the check byte and the number byte.
If a number byte is set in the request frame (request covering more addresses),
the data format in the answer frame is set to 04Hex (unsigned char). The number
byte of the answer frame receives the value of the number byte of the request
frame. A data format eventually entered in the request frame is ignored.
If the additional information of a destination address is requested, the check
byte must include the identification for this additional information. The data
format and the number byte are ignored. The check byte and the
answer byte are set according to the reference list. Valid request frames are
answered with an answer frame. Invalid request frames are answered with
“negative confirmation.”
Example: Requesting the contents of destination address 02H from slave 07H:
00H 00H 00H 10H 07H 02H 00H 00H 19H 16H
—————————————————————————
SYN
SB
SA ZA KB AB FC
EB
Summary of Contents for Wallace & Tiernan Depolox 3 plus
Page 11: ...WT 050 560 001 UA IM 0714 Depolox 3 plus RESIDUAL ANALYZER EVOQUA W3T140188 ...
Page 12: ......
Page 19: ......
Page 39: ...WT 050 560 001 UA IM 0714 26 Depolox 3 plus RESIDUAL ANALYZER EVOQUA W3T140188 ...
Page 40: ......
Page 109: ......
Page 127: ...WT 050 560 001 UA IM 0714 112 Depolox 3 plus RESIDUAL ANALYZER EVOQUA W3T140188 ...
Page 128: ......
Page 146: ...WT 050 560 001 UA IM 0714 130 Depolox 3 plus RESIDUAL ANALYZER EVOQUA W3T140188 ...
Page 147: ......
Page 157: ......