WT.050.560.001.UA.IM.0714
45
Depolox
®
3 plus ReSIDUAl ANAlYZeR
eVoQUA
W3T140188
In order to provide general access, select “Unlocked”. This means that changes
are possible at any time without having to enter a code. Code is no longer dis-
played in the main menu. However, another code number can now be entered
and access denied to all those operators who do not know the new number.
The unit is delivered with the “Lock code set” = 0, “Lock setting” = Unlocked.
If the code number is forgotten enter the back entry code 911.
You can only alter the values in the protected menus once you have set the
correct code number in the SETUP menu and thus proved that you are an
authorized operator.
3.1.5
High - Low Alarms
The Depolox® 3
plus
has independent alarms that can be allocated to the
measured values for chlorine (and optionally for pH or fluoride).
Menu Setting
Alarm Condition
Action
High
X
alarm value
alarm message on
X
(alarm value - deadband)
alarm message off
Low
X
alarm value
alarm message on
X
(alarm value - deadband)
alarm message off
The deadband can be set in the range of one to 25 digits.
Figure 3.4 - High / Low Alarms
Summary of Contents for Wallace & Tiernan Depolox 3 plus
Page 11: ...WT 050 560 001 UA IM 0714 Depolox 3 plus RESIDUAL ANALYZER EVOQUA W3T140188 ...
Page 12: ......
Page 19: ......
Page 39: ...WT 050 560 001 UA IM 0714 26 Depolox 3 plus RESIDUAL ANALYZER EVOQUA W3T140188 ...
Page 40: ......
Page 109: ......
Page 127: ...WT 050 560 001 UA IM 0714 112 Depolox 3 plus RESIDUAL ANALYZER EVOQUA W3T140188 ...
Page 128: ......
Page 146: ...WT 050 560 001 UA IM 0714 130 Depolox 3 plus RESIDUAL ANALYZER EVOQUA W3T140188 ...
Page 147: ......
Page 157: ......