Depolox
®
3 plus ReSIDUAl ANAlYZeR
eVoQUA
W3T140188
4.4.2
Maintenance of the Membrane Probes
NOTE: Do not touch the membrane or reference electrode.
Before unscrewing the membrane cap, remove the elastomer seal to allow air
into the vent hole. Unscrew the membrane cap, clean the reference electrode
with distilled water and then dry with a clean paper towel. Cover the vent hole
with the elastomer seal. Refill the cap with new electrolyte (refer to paragraph
2.3, Membrane Sensor). If the signal is still low or unstable, it may be neces-
sary to replace the cap with a new one (refer to Section 6 - Spare Parts List).
After this procedure, the probe requires a run-in time of about one to two
hours before recalibration.
4.4.3
Troubleshooting
The following troubleshooting table is provided for determining and correcting
most common troubles.
Table 4.4 - Troubleshooting - Membrane Sensor Kit
FAULT CONDITION
POSSIBLE CAUSE
CORRECTIVE ACTION
Insufficient flow of measuring
water
free runoff prevented; back
pressure too high
check for free drain or return
dirt in cell
clean measuring cell
air bubbles
check sample water take-off
screen in water inlet blocked
clean
inlet valve closed
open valve
Air bubbles
leaking O-rings
replace O-rings
incorrect fittings
check and if necessary tighten or
replace
DIS Electrode current too low
electrolyte exhausted
perform sensor maintenance
membrane defective
replace membrane cap
DIS Span cannot be adjusted
measuring cell gives insufficient
current
perform sensor maintenance
The minimum current difference
between un-chlorinated and
chlorinated water with 1 mg/l
residual DIS must be at least >2µA.
If the difference is smaller, carry out
maintenance.
Strong fluctuations in display,
though chlorine concentration
is constant
sample water flow too low
check; if necessary, clean inlet
electrolyte exhausted
perform sensor maintenance
Summary of Contents for Wallace & Tiernan Depolox 3 plus
Page 11: ...WT 050 560 001 UA IM 0714 Depolox 3 plus RESIDUAL ANALYZER EVOQUA W3T140188 ...
Page 12: ......
Page 19: ......
Page 39: ...WT 050 560 001 UA IM 0714 26 Depolox 3 plus RESIDUAL ANALYZER EVOQUA W3T140188 ...
Page 40: ......
Page 109: ......
Page 127: ...WT 050 560 001 UA IM 0714 112 Depolox 3 plus RESIDUAL ANALYZER EVOQUA W3T140188 ...
Page 128: ......
Page 146: ...WT 050 560 001 UA IM 0714 130 Depolox 3 plus RESIDUAL ANALYZER EVOQUA W3T140188 ...
Page 147: ......
Page 157: ......