WT.050.560.001.UA.IM.0714
2
Depolox
®
3 plus ReSIDUAl ANAlYZeR
eVoQUA
W3T140188
1.1
Depolox® 3
plus
Electronics
Power supply
115 V ± 10%, 50 - 60 Hz, 14 VA
F1 Fuse T630 mA, 5 x 20 mm
or
230 V ± 10%, 50 - 60 Hz, 14 VA
F1 Fuse T630 mA, 5 x 20 mm
Input A: Disinfectant
measuring ranges
Free and Total Chlorine
0.20 / 0.50 / 1.00 / 2.00 / 5.00 / 10.0 /
20.0 mg/l
Input A: Chlorine Dioxide
ranges
0.20 / 0.50 / 1.00 / 2.00 / 5.00 / 10.0 /
20.0 mg/l
Input A: Ozone ranges
0.20 / 0.50 / 1.00 / 2.00 / 5.00 / 10.0 /
20.0 mg/l
Input A: Sample water tem-
perature range
Depolox® 3 plus and Depolox® 5
Bare Electrode
41 to 122°F (5 to 50°C)
Membrane Probes
41 to 113°F (5 to 45°C)
Input B: pH measuring ranges
(Dual sensor)
pH 4.00 to pH 10.00, pH 0 to pH 14.00
(Effective range for internal pH compensa-
tion pH 6.0 to pH 8.75)
Input B: Fluoride measuring
range (Dual sensor)
0.20 to 2.00 mg/l
Measurement inputs
1 x disinfection sensor
1 x temperature
1 x pH or fluoride sensor (Dual Input only)
electrically isolated for up to 50 V to
ground
mA outputs 4 - 20 mA
1x for disinfection
1 x temperature
1x for pH or fluoride (with Dual Sensor)
max. 1000 ohm load, electrically isolated
for up to 50 V to ground
Relay outputs
2 freely configurable alarm contacts for
disinfectant signal
2 for ph or fluoride (with Dual Sensor)
Electrically isolated up to 500 V to ground
Relay Contact Rating
5A 1/6 HP 125, 250 V AC or
5A 30 V DC 30 W max.
Interference suppression via suppressor
diodes.
Summary of Contents for Wallace & Tiernan Depolox 3 plus
Page 11: ...WT 050 560 001 UA IM 0714 Depolox 3 plus RESIDUAL ANALYZER EVOQUA W3T140188 ...
Page 12: ......
Page 19: ......
Page 39: ...WT 050 560 001 UA IM 0714 26 Depolox 3 plus RESIDUAL ANALYZER EVOQUA W3T140188 ...
Page 40: ......
Page 109: ......
Page 127: ...WT 050 560 001 UA IM 0714 112 Depolox 3 plus RESIDUAL ANALYZER EVOQUA W3T140188 ...
Page 128: ......
Page 146: ...WT 050 560 001 UA IM 0714 130 Depolox 3 plus RESIDUAL ANALYZER EVOQUA W3T140188 ...
Page 147: ......
Page 157: ......