Manual instruction MM850482 ed.04
5
1.1 Ordering
Code
UniGas 1000 / BTU 1000 7820 – A – B – C – D – E – F
Each Instrument
is
equipped standard
with differential pressure and draft sensor, Tc gas temperature input, Pt100
combustion air temperature input, IR printer port, real-time clock capabilities, rechargeable battery pack, battery charger and
supplied with a Report of Calibration and an instruction manual.
Table A
Sensor n.1
1
O
2
(0-25%)
Table B
Sensor n.2
0
None
2
CO (0-4000 ppm), EN50379-3
2H
CO (0-8000 ppm) H
2
compensated, EN50379-2
Table C
Flue gas probe (line filter and water trap included)
1
180mm flue gas probe or draft (single hose) BB610047
2
300mm flue gas probe or draft (single hose) BB610048
Table D
Option
0
None
P
Built-in impact printer
Table E
Mains adapter / charger
1
115V ±10% 50/60Hz - USA plug
2
230V ±10% 50/60Hz - Schuko plug
3
230V ±10% 50/60Hz - UK plug
4
230V ±10% 50/60Hz - European plug
5
100V ±10% 50/60Hz - USA/Japan plug
Table F
Calibration certificate
1
Eurotron
report
1.2 Specifications
•
Type
: 1 or 2 cells Hand-held flue gas analyzer.
•
Calibration
: automatic calibration procedure at 60 seconds with smart autozero switch-On.
•
Self-diagnosis
: Sensors efficiency test with diagnostic messages display.
•
Fuel types
: Up to 10 selectable from keyboard (most common preprogrammed).
•
Pump
: rate of flow 0.8 lit / head –70mbar - All data measured using 3 meters long probe and line filter
connected. Load loss = 10 mbar using a 3 meters probe extension
•
Power supply
: High capacity Ni-MH rechargeable battery pack / external battery charger.
•
Charging time :
8h at 90% with instrument off.
•
Battery life
: 6h continuous operation (without printing and back-light).
•
Printer power supply
: from the analyser battery pack.
•
Printer Report Header
: 4 programmable lines.
•
Display
: 40x58 mm alpha-numeric LCD with backlight device.
•
Infrared port:
compatible with HP82240B wireless printer
•
Smoke measurement
: Using the optional external manual pump.Index memory store and printout
capability as standard.
•
Flue gas Probe
: stainless steel with rubber handle and incorporated temperature sensor. 2 meters long
rubber hose included.
•
Water trap
: external with purge plug and filter.
•
Line filter
: with replaceable cartridge.
•
Working temperature
: from –5°C to +45 °C (up to 50°C for short time)
•
Storage temperature
: from –20 to +60°C (3 months maximum at temperatures exceeding the operational
limits).