Manual instruction MM850482 ed.04
27
•
Press the [
] key to print the test report.
•
Press the [
ESC
] key to return the measuring page.
6.2.3 Pressure
Kit
EE300248
This pressure kit has been designed to perform the leak
tests.
Individual elements (contents of case):
•
n° 2 conical gas test plugs with nipple to connect the adapters to the gas network
•
n° 1 hand bellow for pressurize the gas networks
•
n°1 ‘T’ adapter with gas cut-off valve
•
n°1 connector for the analyser input
6.3 Smoke
Measurement
To avoid condensation inside the smoke pump, the pump should aspirate ambient air before the gas sample.
This can be obtained aspirating the air by the fireside (not in the exhaust gas connection pipe) without using
the paper filter.
Now introduce the filter in the pump.
LEGEND
P1-P4:
pressure measurements
Drop:
pressure drop after the test end (P4-P1)
Duration :
Test Time