Manual instruction MM850482 ed.04
29
•
Press
[
ENTER
] and input the smoke index values into the following page:
•
Change the 1st test value by pressing the [ ] and [ ] keys. Press the [
ENTER
] key to store the value and
input the test 2 value.
•
Change the 2nd test value by pressing the [ ] and [ ] keys. Press the [
ENTER
] key to store the value
and input the test 3 value.
•
Change the 3rd test value by pressing the [ ] and [ ] keys. Press the [
ENTER
] key to store the value
NOTE:
IF YOU INPUT A WRONG VALUE
,
PRESS THE
[
ESC
]
KEY TO RETURN BACK TO THE MAIN MEASURING PAGE AND
REPEAT THE PRCEDURE
.
The following page will appear:
If during the smoke measurements you notice some petroleum derivatives on the filter paper, you have to
input "YES" on the "der:" field by using the [
] key.
Press the [
ENTER
] key to store the value.
Press
[
ESC
] to return to the measure page.
6.4
Analysis Data Print
UniGas 1000
has two different possibilities to print the flue gas analysis data. You can use the optional
internal impact printer or the standard IR port for your HP 82240B wireless thermal printer.
6.4.1 Built-In
Printer
(option)
You can order the
UniGas 1000
complete with an internal 24-column impact printer. The printer is integrated
with the instrument and the power is supplied by the same batteries as the analyzer. The impact model gives
you one more advantage: more readability and long shelf life of your documents.
To print the data analysis follow the procedure:
Wait for the stabilization of the measurenment on the display.
Press the [
] key.
The display shows the last setting for number of copies to be
printed and the set document format.
Check that the internal printer is enabled. Press the [
MENU
] key
and verify that it displays the “
PRINTINT
” message.
If you do not need any changes press the [
] key again. The
instrument will print the report as programmed
If you need any changes, read the procedure on paragraph 6.3.3.