Manual instruction MM850482 ed.04
13
4 RECOMMENDATIONS
The analyzer should be used in environments where the temperature is between -5°C and +45°C.
When the analysis are completed, before switch the instrument off, remove the probe from the gas
pathway and wait about 30 seconds cleaning the pneumatic circuit from gas.
Do not use the analyzer with clogged filters or filled with humidity.
Before placing the probe in the case, be sure that the probe, the water trap and the hose are clean and
cold.
If you have ordered a double hose sampling probe, insert both connectors to the instrument during the
gas analysis (pay attention to colour correspondence on hose plugs).
To empty the water trap from the condensate water, remove the draining plug. DO NOT OPEN the water
trap. DO NOT leave the water trap without the draining rubber plug (cat. EE650081).
For the best efficiency and accuracy, we suggest calibrating the instrument every year
.
WARNING
I
F THE
U
NI
G
AS IS STORED AT TEMPERATURE EXCEEDING THE OPERATIVE LIMITS
,
THE ANALYZER NEEDS SOME MINUTES
TO WARM
-
UP TO THE ACTUAL AMBIENT TEMPERATURE
,
BEFORE STARTING THE OPERATION
.
4.1 Power
supply
UniGas
can be powered from:
internal rechargeable Ni-MH battery.
External battery charger, supplied as a standard accessory (the batteries must be installed).
Optional external DC auto battery charger (the batteries must be installed).
The Ni-MH rechargeable batteries allows a long operation time and do not need maintenance. The same
batteries power both the instrument and the internal printer. The internal battery will grant 6 hours continuous
operation (without printing and display back light off).
N
OTE
:
90%
BATTERY CHARGING CAN BE OBTAINED IN
8
HOURS WITH THE
U
NI
G
AS SWITCHED OFF
.
During operation a fully battery symbol “
≠
” will be displayed on the display. This symbol means that the
batteries are completely full. When the batteries will be discharged the symbol “
–
” will appear and the
instrument still has about 20 minutes operation capability to end the running analysis.
The battery symbol indicates that a full charge is required. Use only the dedicated battery charger supplied by
Eurotron together the instrument.
CAUTION:
O
LD BATTERIES CAN LEAK AND CAUSE CORROSION
.
N
EVER LEAVE RUN DOWN BATTERIES IN THE
INSTRUMENT
WARNING
T
HE INSTRUMENT IS SHIPPED WITH AN AVERAGE LEVEL OF BATTERY CHARGE
.
A
FTER UNPACKING
,
A FULL CHARGE OF THE
BATTERY IS RECOMMENDED
,
BY CONNECTING THE INSTRUMENT TO THE MAIN LINE THROUGH THE BATTERY CHARGER
(OFF
CONDITION
)
FOR
8-10
HOURS AT A TEMPERATURE BETWEEN
10
AND
30°C