Manual instruction MM850482 ed.04
6
•
Dimensions
: 115x90x330 mm
•
Weight
: 1.1 kg (battery and printer included)
•
Warranty:
2 years for sensors, connectors, battery, printer and pump.
Accuracies and Ranges
Parameter
Sensor Type
Range
Resol.
Max
Response
Accuracy
O
2
Electrochemical
from 0 to 25.0%
0.1%
20 sec.
±
0.2% Vol.
CO
Electrochemical
from 0 to 4000 ppm
1 ppm
50 sec.
±
10 ppm < 300ppm
±
4% rdg. up to 4000 ppm
CO
H
2
compensated
Electrochemical
from 0 to 8000 ppm
1 ppm
50 sec.
±
10 ppm < 300ppm
±
4% rdg. up to 2000 ppm
±10% rdg. elsewhere
CO
2
Calculated
from 0 to 99.9 %
0.1 %
Tair
Pt100
from -10 to 100 °C
0.1 °C
±
0.5 °C
Tgas
Tc K
from 0 to 600.0 °C
0.1 °C
±
1 °C
∆
T
Calculated
Pressure / draft
Piezoresistive
from -10 to 120 hPa
0.01 hPa
±
3Pa <300 Pa
±1% rdg. elsewhere
Excess air
Calculated
from 1.00 to infinite
0.01
Efficiency
Calculated
from 0 to 120%
0.1 %
(also for condensing boilers with automatic detection)
Smoke index
External pump
from 0 to 9
All emission measurements can be displayed with reference to a programmable O2 value.
The relative accuracy shown are expressed as absolute or % of rdg errors at -5°C to +40°C ambient temperature.
The maximum response time shown is referred to 90% signal changes.
The pressure relative accuracy shown is valid only after the auto-zero procedure
Accuracy limits are stated as % of reading. An additional ±1 digit error has to be considered.
The stated pressure relative accuracy is valid only after the zero procedure.
Measuring reading can be directly converted from °C to °F, ppm to mg/Nm3 and from hPa to mmH2O, mbar, inH2O.
Specifications may change without notice.
Fuel technical data
The instrument includes as standard the technical data for 4 of the most common fuels.
Up to 10 fuel types can be set up by factory on demand.
1.3 Calibration
certificate
Each
UniGas 1000
portable gas analyzer is factory calibrated and certified against
Eurotron
Standards that
are periodically certified by an International recognized Laboratory, and shipped with a Report of Calibration
stating the nominal and actual values and the deviation errors.
1.4 Electromagnetic
compatibility
The instrument case, made in shock-resistant injection molded ABS + polycarbonate has designed to fulfill the
directive 89/336/CEE Electromagnetic Compatibility. See Appendix A1 for EMC declaration of conformity.