Manual instruction MM850482 ed.04
32
6.5
Carbon Monoxide Safety Room Test
A procedure can be selected to monitor the CO in ambient air using the internal sensors. An internal program
allows the CO max measurement in atmospheric boiler room test with 15 values logging. An external optional
probe is also available for continuous surveillance of the ambient and operator safety. Both acoustic and
visual alarm are available.
Remove any pneumatic connection from the
UniGas 1000
gas INLET.
Press the [
MENU
] key until the following page will be displayed:
Press the [
ENTER
] key to monitor ambient CO.
The test will detect and store the CO value for 15 times, one time per minute. The complete test time is 15
minutes. The instrument will display a 900 seconds countdown and the number of the already stored values
on the third row on the display as shown on the page below:
NOTE:
A
MBIENT
CO
MONITORING IS ACTIVE ONLY ON
“COAMBINT”
OPERATIVE PAGE
.
Press the [
] key at the end of the test to print the report