100/180 MM GRAPHICS RECORDER: USER GUIDE
HA261742
Issue 16 Nov 04
User Guide
Page 26
3.3 FIRST SWITCH-ON
When power is applied the recorder initialises, and once this process is complete, the home page is displayed. It is
unlikely that this will contain any useful information because the input channels will not, as yet, have been configured
to suit the type of input signals being applied to them, as described in section 4.
Notes:
1
There is no on/off switch associated with the recorder.
2
Date, time and the message 'Power Up' are printed on the chart each time power is applied to the re-
corder, followed (if the Auditor Pack option is fitted) by a similar message giving 'Config Revision' and
'Security Revision'.
3
A red line is drawn across the width of the chart at power up
The recorder has four security levels as follows:
Logged out
Initially, no access to recorder configuration is possible. Only Archive, Login/security and the System
'About' functions can be accessed - via the root menu. Limited or full access can be permitted from
'Engineer' level.
Operator*
No access to recorder configuration is possible until access permissions have been set up.
describes how limited or full access can be permitted by an operator with 'Engineer' level access.
Engineer*
Accessed initially, by entering '10' as the password (
). Full access to all recorder
functions is available.
describes how the Engineer level password can be edited and an
Operator password can be entered if required. The section also describes how access permission to
some or all of the recorder functions can be granted, or not, to individual user names and default secu-
rity levels (except 'Service').
Service
Full access to all recorder functions and to areas of recorder memory for diagnostic purposes. For use
only by Service Engineers.
*Note: For units fitted with software versions 3.1 onwards, AND which have the Auditor Pack Option
enabled, the default Engineer and Operator passwords are both '100'.