Model 2090 Multi-Device Controller
download from the ETS-Lindgren web site at
www.ets-
lindgren.com
.
MB-V IDENTIFICATION
The MB-V interface can be identified visually by the
control board enclosure mounted on the outside of the
motorbase with two (2) metal fiber-optic connectors
extending parallel to the main enclosure. From the Model
2090, use the new BASE parameter in the parameter menu
to verify that the motor base has a flash upgradeable MB-
IV board and to determine the existing firmware revision.
WARNING: Under no circumstances should a motor base
with firmware revision less than V1.04 be updated.
Firmware versions earlier than V1.04 indicate an earlier
revision interface board which is incompatible with
firmware revisions V1.04 or later. If you determine you
have an MB-IV motor base with an earlier revision, contact
ETS for more information.
NOTE to 2088 and 5901 users: Upgrading older 2088 or
5901 motor bases will likely cause a change in
performance, slightly increasing the maximum speed. This
increase in speed may also affect positioning accuracy. If
you encounter difficulties after upgrading, contact ETS for
additional information on possible adjustments to your
2088 or 5901 motor base.