Model 2090 Multi-Device Controller
ERE
Compatibility:
2090 (Firmware Rev 2.3)
Description:
Set Device Dependent Error Enable Register. Causes changes in the contents
of the Device Dependent Error Enable Register. Requires a decimal
argument in binary form to set the register. The bits of the argument
correspond to the bits of the Device Dependent Error Register. Setting a bit
in this register allows the corresponding bit in the Device Dependent Error
Register to cause the DDE bit in the Status Byte Register to be set.
Devices:
Towers, Turntables
Syntax:
ERE XXXXX
XXXXX
Integer value representing a sixteen bit binary
number 0-65535
Example:
CALL ibwrt (TowerUD%, "ERE 64")
' Allow a polarization violation
' to set the DDE bit of STB
See Also:
ERE?, ERR?, *STB?
ERE?
Compatibility:
2090 (Firmware Rev 2.3)
Description:
Query Device Dependent Error Enable Register. Prepares the controller to
respond with the contents of the Device Dependent Error Enable Register
when queried. Value returned is an integer number in the range of 0 to
65535. This value, when expressed in binary form, represents bit values of
the Device Dependent Error Enable Register.
Devices:
Towers, Turntables
Syntax:
ERE?
Response:
<ERE
bits>
Integer value representing a sixteen bit binary
number 0-65535.
Example:
CALL ibwrt (TableUD%, "ERE?")
' Query ERE register
CALL ibrd (TableUD%, rdbuf$)
' Read response
See Also:
ERE, ERR?, *STB?
©ETS-Lindgren, April 2006
111
Revision G– P#399199