Appendix B: Maintenance
Model 2090 Multi-Device Controller
(out-to-in and in-to-out) and that the device(s) are powered
on.
If none of these options corrects the problem, it may also be
ecessary to insure that no cables inside the Model 2090
V3.11 of the Model 2090 firmware contains additional
take
advantage of all of these enhancements, the variable speed
n
were disconnected or knocked loose during the installation.
If the problem still persists, try replacing the original
EPROMs and jumper settings to verify the Model 2090 still
works normally. The SBC may not be compatible with the
new firmware. Contact ETS for additional assistance.
VARIABLE SPEED DEVICES
enhancements for variable speed devices. To
motor base (MB-IV) firmware revision must be at V1.10a
or later. Use the new BASE parameter in the parameter
menu to determine the firmware revision (and to verify that
the motor base has a flash upgradeable MB-IV board). Go
to
www.ets-lindgren.com
or contact ETS-Lindgren to
obtain the motor base flash upgrade.
WARNING: Under no circumstances should a motor base
with firmware revision less than V1.04 be updated.
Firmware versions earlier than V1.04 indicate an earlier
revision interface board which is incompatible with
firmware revisions V1.04 or later. If you determine you
have an MB-IV motor base with an earlier revision, contact
ETS for more information.
146
© ETS-Lindgren– April, 2006
Revision G– P# 399199
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com