Mini MIRI
®
Dry and Mini MIRI
®
Humidity Multiroom IVF incubators User Manual
Rev. 3.0
74
For safety reasons, this unit has a built-in digital gas pressure sensor control that moni-
tors the incoming gas pressure and alerts the user if any drop is detected.
It is recommended to check the CO
2
gas pressure in the menu by inspecting the value for
an item called
‘CO2
P’
(CO
2
pressure).
PASS
: The value must be 0.4
–
0.6 bar.
Please refer to the
“18.1
CO
2
gas
pressure”
section for more information.
36.5 N
2
gas pressure check
The Mini MIRI
®
Dry or Mini MIRI
®
Humidity multiroom IVF incubator requires a pressure
of 0.4
–
0.6 bar on the input N
2
gas line. This gas pressure must be held stable at any time.
For safety reasons, this unit has a built-in digital gas pressure sensor control that moni-
tors the incoming gas pressure and alerts the user if any drop is detected.
It is recommended to check the N
2
gas pressure in the menu by inspecting the value for
an item called
‘N2
P’
(N
2
pressure).
PASS
: The value must be 0.4
–
0.6 bar.
Please refer to the
“18.2
N
2
gas
pressure”
section for more information.
37 The Maintenance guide
Your Mini MIRI
®
Dry or Mini MIRI
®
Humidity multiroom IVF incubator from Esco Medical
contains high precision quality components. These components are chosen to ensure the
high durability and performance of the equipment.
However, continual validation of the performance is necessary.
User validation should be done as a minimum according to instructions given in the "33
The Validation guide" section.
If problems are encountered, contact Esco Medical or your local representative.
However, to sustain the high-performance level and avoid system errors, the owner is
responsible for having a certified technician who performs components replacements ac-
cording to table 37.1.