Mini MIRI
®
Dry and Mini MIRI
®
Humidity Multiroom IVF incubators User Manual
Rev. 3.0
53
Mini MIRI
®
Dry and Mini MIRI
®
Humidity multiroom IVF incubators are used. The
manufacturer recommends that the period between validation should be no
longer than 14 days.
2.
VOC/HEPA filters must be replaced every 3 months.
3.
In-line HEPA filters must be replaced yearly during annual maintenance.
4.
According to the clinical practice intervals, suitable cleaning is in the laboratory
where the Mini MIRI
®
Dry and Mini MIRI
®
Humidity multiroom IVF incubators are
used. The manufacturer does not recommend periods longer than 14 days be-
tween cleaning.
It is essential to perform the inspection and service at the intervals indicated
in the MAINTENANCE section below. Failure to do so can have serious adverse out-
comes, causing the unit to stop functioning as expected and cause damage to sam-
ples, patients or users.
Warranty is considered to be void if service and maintenance are not fol-
lowed.
Warranty is considered void if service and maintenance procedures are done
not by trained and authorized personnel.
29 Emergency Procedures
Total loss of power to or on the unit:
•
Remove all the samples and place them in an alternative or backup device that is
not affected by the problem.
•
Without the power source, the internal temperature of the Mini MIRI
®
Dry and
Mini MIRI
®
Humidity multiroom IVF incubator will drop below 35 °C
after being
10 minutes in an ambient environment of 20 °C.
•
The CO
2
concentration will remain within 1% of the setpoint for 30 minutes if the
lids remain closed.
•
If a longer time to turn the power back on is needed, it may be useful to cover the
unit with insulating blankets to slow the temperature drop.
If a single temperature alarm goes off:
•
Remove the samples from the affected compartment. They can be relocated to any
of the other compartments, which happens to be unoccupied. All compartments
are separate so that the remaining ones will function normally.