Mini MIRI
®
Dry and Mini MIRI
®
Humidity Multiroom IVF incubators User Manual
Rev. 3.0
28
13.4.2 CO
2
sub-menu
Press the rotatory button on the CO
2
menu to enter the CO
2
sub-menu.
The first item in the CO
2
sub-menu is CO
2
sensor calibration:
Calibrate CO
2
by pressing the rotatory button and rotating it (
↻
) or (
↺
) to adjust.
Move to the next CO
2
sub-menu item with (
↻
) rotation or one step up with (
↺
) rotation.
Toggle CO
2
regulation on/off by pressing the rotatory button and rotating it (
↻
) or (
↺
).
The default status for the CO
2
control is OFF.
Move to the next CO
2
sub-menu item with (
↻
) rotation or one step up with (
↺
) rotation.
CO
2
flow rate is shown (it cannot be adjusted):
It shows the amount of CO
2
gas put into the system while regulating. The volume is shown
in liters/hour. It usually will fluctuate along with the CO
2
regulation.
Move to the next CO
2
sub-menu item with (
↻
) rotation or one step up with (
↺
) rotation.
CO
2
internal pressure rate is shown (it cannot be adjusted on the Mini MIRI
®
Dry and Mini
MIRI
®
Humidity multiroom IVF incubator. It is adjusted on the external gas regulator):