Mini MIRI
®
Dry and Mini MIRI
®
Humidity Multiroom IVF incubators User Manual
Rev. 3.0
34
The Mini MIRI
®
Dry and Mini MIRI
®
Humidity multiroom IVF incubators must be set to
premix gas work mode.
Please follow these instructions
:
Press the rotatory button
–
enter the menu.
The user can exit the menu by rotating the rotatory button (
↺
).
Press the rotatory button on the service menu to enter the service sub-menu.
The display will show the currently installed firmware version. Move to the next service
sub-menu item with (
↻
) rotation or one step up with (
↺
) rotation.
The display will show the
“GAS”
function:
Press the rotatory button to enter the service sub-menu and rotate (
↻
) or (
↺
) to choose
“PREMIX”.
Press the rotatory button to select the premixed gas work mode.
Then exit the menu by (
↺
) or press the rotatory button and hold it until the main menu
does not appear.
The CO
2
setpoint must be 0.1% lower than the premixed gas measured value
(i.e., 4.9% if 5.0% measured).
The O
2
setpoint must be 1% higher than the premixed gas measured value gas
(i.e., 6.0% if 5.0% measured).
For changing the CO
2
and O
2
setpoints, please read the 16.2 and 16.3 sections in this man-
ual.