Mini MIRI
®
Dry and Mini MIRI
®
Humidity Multiroom IVF incubators User Manual
Rev. 3.0
39
"P" stands for pressure.
The display will lock on the alarm condition and will stop alternating between the stand-
ard status messages. If the mute key is pressed, the display will shift to normal status and
show the parameters for 5 minutes until the audio alarm comes back on again. The mute
alarm key will still show the alarm condition by blinking red while the alarm is muted.
15.4 Multiple alarms
When there are two or more alarms, the display will indicate this by showing first "A
MULTI" and then the alarm conditions:
The flow will be forced according to the alarms. The temperature alarms have 1
st
priority,
gas level alarms 2
nd
, and gas pressure 3
rd
.
15.5 Alarm UVC light (applicable only for the Mini MIRI
®
Dry model)
Alarms on UV-C light will show only as a warning message during the normal status. A
red "S" will appear.
There will be no audio alarm.
The user should consult the distributor for further guidance or service inspection. Only
when the UV-C light works again will the "S" disappear.
15.6 Loss of power alarm
If the power is disconnected, the Mini MIRI
®
Dry and Mini MIRI
®
Humidity multiroom
IVF incubator will give an audio alarm for approximately 4 seconds, and the LED in the
mute alarm key will flash.
Figure 15.2
Alarm key which indicates the alarm condition