Mini MIRI
®
Dry and Mini MIRI
®
Humidity Multiroom IVF incubators User Manual
Rev. 3.0
41
16.3 The O
2
gas concentration set point
The O
2
concentration can be adjusted in the range between 3.9% to 19.9%.
The default O
2
setpoint is 5.0%
To change the O
2
gas concentration setpoint, follow these instructions:
1.
When the display shows the O
2
gas concentration:
2.
Press the rotatory button and rotate (
↻
) or (
↺
) to adjust the setpoint.
3.
After changing the value, press the rotatory key once more to save the value.
If the display does not show the current temperature reading, rotatory button rotation
(
↻
) or (
↺
) will toggle between the temperature, CO
2
, O
2
and mode readings.
16.4 The culture mode
The culture mode can be set for under
“oil
culture”
or
“open
culture”.
“Under
oil”
culture
mode is used when the culture media has an oil or Paraffin overlay.
“Open
culture”
mode
is used when the culture media does not have any overlay.
The default setting is
“O
il culture
”
mode.
To change the culture mode, follow these instructions:
1.
When the display shows the culture mode:
2.
Press the rotatory button and rotate (
↻
) or (
↺
) to change the mode.
3.
When the display shows the desired/correct mode, press the rotatory button. The
mode is now set.
If the display does not show the current mode reading, rotatory button rotation (
↻
) or
(
↺
) will toggle between the temperature, CO
2
, O
2
and mode readings.