Mini MIRI
®
Dry and Mini MIRI
®
Humidity Multiroom IVF incubators User Manual
Rev. 3.0
31
Rotate rotator button (
↻
) to adjust the O
2
calibration to the desired level by (
↻
) or (
↺
)
rotation. In this case, we want to change to 5.3%. Rotate (
↻
) till the display shows 5.0,
5.1, 5.2 and 5.3. When O
2
equals measured O
2,
press the rotatory button. The new value
is stored and the O
2
sensor calibration has been modified.
Calibration is performed by adjusting the O
2
level according to the gas sam-
pling outlet's measurement by an external reliable O
2
measurement device.
Calibration values should only be changed by a trained user or the technician,
according to specific measurements. Done with the calibrated device.
13.4.4 UVC light sub-menu (applicable only for the Mini MIRI
®
model)
Press the rotatory button on the UV-C menu to enter the UV-C light sub-menu.
Toggle UV-C light regulation on/off by pressing the rotatory button and rotating it (
↻
) or
(
↺
).
The default status for the UV-C light is
“ON”
.
The UV light will automatically switch off when the unit is turned off.
For optimal air cleaning, it is recommended to have the UV-C light set to
“ON”
when the unit is used.