![Esco Medical Mini MIRI Dry User Manual Download Page 71](http://html1.mh-extra.com/html/esco-medical/mini-miri-dry/mini-miri-dry_user-manual_2426159071.webp)
Mini MIRI
®
Dry and Mini MIRI
®
Humidity Multiroom IVF incubators User Manual
Rev. 3.0
71
•
Attach the power again and switch-on the Mini MIRI
®
Dry or Mini MIRI
®
Humidity
multiroom IVF incubator.
•
Let the Mini MIRI
®
Dry or Mini MIRI
®
Humidity multiroom IVF incubator run
empty for at least 20 minutes before inserting a sample.
35.12 Test documentation form
The "Installation report" form must be completed with the tests-passed status
filled by installation personnel and submitted to Esco Medical before the device is
taken into clinical use.
35.13 Recommended additional testing
35.13.1 A VOC meter (applicable only for the Mini MIRI
®
Dry model)
A sample should be taken just above the Mini MIRI
®
Dry or Mini MIRI
®
Humidity multi-
room IVF incubator with the VOC meter. The reading should be noted down as the back-
ground VOC level. Then a sample is taken from the gas sample port number
–
2.
Pass: 0.0 ppm VOC.
Ensure that the sample lines do not contain any VOC.
35.13.2 A laser particle counter
A sample should be taken just above the Mini MIRI
®
Dry or Mini MIRI
®
Humidity multi-
room IVF incubator with the laser particle counter. The reading should be noted down as
the background particle level. Then a sample is taken from the gas sample port number
–
2.
Pass: 0.3-micron < 100 ppm.
Ensure that the sample lines do not contain any particles.
36 Clinical use
Congratulations! Your device is now ready for clinical use with the validation tests com-
pleted and the test report submitted to Esco Medical.
It should provide many years of stable service.
It is necessary to monitor the performance of the device continually.
Use the below scheme for in-use validation.