52
Figure 26.2
"Open circuit" alarm mode
Whenever the Mini MIRI
®
Dry or Mini MIRI
®
Humidity
multiroom IVF incuba-
tor's power cord is disconnected from the power source, this switch will automati-
cally indicate an alarm! It is an extra safety feature intended to alert the personnel
in case of a power cut in the laboratory.
27 Writing area on the compartment lids
Each compartment lid on the Mini MIRI
®
Dry or Mini MIRI
®
Humidity multiroom IVF in-
cubator is made from white glass, optimized for writing text. The compartment's patient
data or content can be noted down for easy reference during the incubation process.
The text can be wiped off with a cloth afterward. Use only a suitable non-toxic pen that
allows the text to be erased later and will not damage the incubated samples.
Figure 27.1
Area for patient information
28 Maintenance
The Mini MIRI
®
Dry and Mini MIRI
®
Humidity multiroom IVF incubators are designed to
be user-friendly. Reliable and safe operation of this equipment is based on the following
conditions:
1.
Correct calibration of temperature and gas level, using high-precision equipment
in the intervals prescribed based on clinical practice at the laboratory, where the