8-4
27
The internal temperature of the unit was
too high. The maximum HF power was
automatically reduced.
Always set up the unit in such a way that air can reach
the housing.
28 to 29 These errors concern functions in the unit Please notify the Technical Service
30
The load resistance of the unit was too
low
Either the contact surface of the active electrode was
too large or the power limitation set too low, or there
was contact between an active electrode and a metallic
instrument, e.g. trocar sheath.
31
The HF generator of the unit was
overloaded too long. The maximum
power output was automatically reduced.
The unit can supply briefly more than 200 watts power.
More than 400 watts, averaged over 1 second, are not
permitted for reasons of safety.
32
The instrument at the monopolar socket
is activated by two switches at the same
time, e.g. by the fingerswitch and
footswitch.
Fingerswitch, footswitch defective.
Always activate instruments with one switch only.
Please inform Technical Service.
33
The instrument at the monopolar socket
is activated by the fingerswitch. During
activation, the instrument is activated by
the footswitch. The fingerswitch is
deactivated. The footswitch remains
activated.
(Reverse also possible)
Fingerswitch, footswitch defective.
Always activate instruments with one switch only.
Please inform Technical Service.
34
The instrument at the bipolar socket is
activated by two footswitches at the same
time.
Footswitch defective.
Always activate instruments with one swith only.
Please inform Technical Service.
35
The instrument at the bipolar socket is
activated (footswitch, Auto Start).
During activation, the instrument is
activated by the footswitch. The
instrument is deactivated. The footswitch
remains activated.
Auto Start, Auto Stop, footswitch
defective
Always activate instruments with one switch only.
Please inform Technical Service.
36
The Auto Start key was pressed while the
coagulation electrode was already
touching the tissue
Always first set the required activation mode and then
apply the coagulation electrode to the tissue to be
coagulated
37
This error concerns functions in the unit
Please notify the Technical Service
38
Short between the two poles of a bipolar
instrument, i.e. a bipolar forceps
Avoid shorts
39 Not
available
40 to 49 This error concerns functions in the unit
Please notify the Technical Service
50 to 51 Errors in a front panel key
Please notify the Technical Service
CAUTION
This error list contains error descriptions relevant only to the operator.
An expanded error list, which particularly describes technical errors, can be found in the technical service
documentation.
Summary of Contents for 10128-002
Page 1: ...ERBOTOM ICC 200 Instruction manual 04 05 ERBE...
Page 2: ......
Page 8: ...1 2...
Page 10: ...2 2...
Page 18: ...3 8...
Page 19: ...International Standard...
Page 20: ......
Page 34: ...4 14...
Page 46: ...5 12 FORCED COAGULATION Power output as a function of the power limitation RL 350 ohms...
Page 56: ...7 2...
Page 62: ...9 2...
Page 64: ...10 2...
Page 66: ...11 2...
Page 72: ...12 6...