8-3
Error
No.
Error
How to proceed
0
No error
1
No HF output voltage
Notify the Technical Service
2
HF output voltage too high
Notify the Technical Service
3
No HF output voltage
Notify the Technical Service
4
HF output voltage too high
Notify the Technical Service
6
Activation error
Notify the Technical Service
7
Activation error
Notify the Technical Service
9
Time limit exceeded
Heed maximum time limit
10
Erroneous setting of the AUTO CUT
function field during activation
Before activation of a cutting mode, the AUTO CUT
function field must be completely set
11
Erroneous setting of the AUTO COAG
function field during activation
Before activation, the AUTO COAG function field
must be completely set
12
Erroneous setting of the AUTO
BIPOLAR function field during
activation
Before activation, the AUTO BIPOLAR function field
must be completely set
13
The contact surface between the neutral
electrode and the patient is too small or
the neutral electrode was not connected
to the unit.
This error is recognized and reported in
the Standby mode or during activation of
the unit.
Only for NESSY Version 2
Error in the connection between the
neutral electrode and unit
Only for NESSY Version 3
Short between the two contact surfaces
of a divided neutral electrode or in the
cable to the neutral electrode
Use a sufficiently large neutral electrode and apply the
entire surface.
Check the connection of the neutral electrode to the
unit.
Check the connection between the neutral electrode and
unit. Probably a faulty cable or faulty plug
Eliminate short or exchange cable
19
The contact surface between the neutral
electrode and the patient was too small
Use a sufficiently large neutral electrode and apply the
entire surface.
20
The neutral electrode was applied in the
incorrect direction
See Chapter 4.3.3
NESSY
21
The neutral electrode was applied in the
incorrect direction
See Chapter 4.3.3
NESSY
22
When switching on the power switch, the
yellow pedal on the footswitch or the
yellow key on the electrode handle was
already pressed
Check whether there was an operator error or whether
the yellow pedal of the footswitch or the yellow key on
the electrode handle is defective
23
When switching on the power switch, the
blue key on the electrode handle
connected to the CUT/COAG socket was
pressed
Check whether there was an operator error or whether
the blue key on the electrode handle or the cable for the
electrode handle is defective
25
When switching on the power switch, the
blue pedal was already pressed
Check whether there was an operator error or whether
the blue pedal is defective
26
When switching on the power switch,
there was an electrically conductive
connection between the two poles of the
bipolar instrument
Always be certain that bipolar or monopolar active
electrodes or instruments are laid onto electrically
nonconductive surfaces
Summary of Contents for 10128-002
Page 1: ...ERBOTOM ICC 200 Instruction manual 04 05 ERBE...
Page 2: ......
Page 8: ...1 2...
Page 10: ...2 2...
Page 18: ...3 8...
Page 19: ...International Standard...
Page 20: ......
Page 34: ...4 14...
Page 46: ...5 12 FORCED COAGULATION Power output as a function of the power limitation RL 350 ohms...
Page 56: ...7 2...
Page 62: ...9 2...
Page 64: ...10 2...
Page 66: ...11 2...
Page 72: ...12 6...