6-6
The unit is equipped with a specific front panel basic setting. The specific basic setting is defined
at the factory and cannot be changed by the user. This specific basic setting only appears if the
basic setting freely programmable by the user is lost due to an error. For the ICC 200 unit, the
following specific basic setting is defined at the factory:
AUTO CUT EFFECT
3
AUTO CUT MAX.WATT
120 watts
ENDO CUT
switched on
T-on time
50 ms
T-off time
750 ms
Monopolar SOFT COAG
60 watts
Monopolar FORCED COAG
60 watts
Monopolar ARGON COAG
60 watts
AUTO COAG BIPOLAR
40 watts, AUTO START switched off