5-2
ARGON PLASMA COAGULATION
Applies only to ERBOTOM ICC 200 with Endo Cut and Argon Coag
HF voltage waveform
pulse-modulated alternating voltage
Crest factor C, at R
L
=500 ohms
C = 8 at 99 W
max
, C=15 at 5 W
max
Rated frequency of the HF voltage
1 MHz
Peak value of the HF voltage
max. 4 kVp
HF rated power
99 watts at 350 ohms
HF power limitation
from 1 watt to 99 watts in 1 watt steps
Setting of the HF power limitation
via up/down keys
Display of the HF power limitation
7-segment display, 2 decimal places
Precision of the power limitation
+/- 1 digit or +/- 15%
Activation
via key or via pedal
HF connecting sockets
1
BIPOLAR COAGULATION / AUTO BIPOLAR
HF voltage waveform
unmodulated sinusoidal alternating voltage
Crest factor C, at R
L
=500 ohms
C = 1.4 for all settings
Rated frequency
330 kHz
Peak value of the HF voltage
max. 190 V
Constancy of the HF voltage
automatically controlled up to P
HFmax.
Dynamic internal impedance
0 ohms in rated range
HF rated power
120 watts at 125 ohms
HF power limitation (P
HFmax.
)
from 1 watt to 120 watts in 1 watt steps
Setting the HF power limitation
via up/down keys in 1 watt steps
Display of the HF power limitation
7-segment display, 3 decimal places
Precision of the HF power limitation
+/- 1 digit or +/-15%
Activation of Bipolar coagulation
via pedal or Auto Start
Auto Start delay
from 0 to 10 s via program no. 23
HF connecting sockets
1
SAFETY FEATURES
Protection class according to EN 60 601-1
I
Type according to EN 60 601-1
CF
Switching of the neutral electrode
Floating Output
Monitoring of single-surface neutral electrodes
Automatic monitoring of the electric connection
between neutral electrode and high-frequency surgical
unit
Monitoring of dual-surface neutral electrodes
Automatic monitoring
a) of the electric connection between the neutral
electrode and high-frequency surgical unit
b) as well as between the neutral electrode and patient
c) of the symmetry of the HF partial currents i
HF1
/
i
HF2
d) of the HF current I
HF
dependent on the contact
resistance R
ü
between the partial surfaces of the
neutral electrode
Max. resistance R
ü
between the partial surfaces of the
divided neutral electrodes
120 ohms
±
20 ohms
Warning signals S
w
dependent on I
HF
and R
ü
see Diagram S
w
= f ( I
HF
, R
ü
)
Monitoring of the HF output parameters
HF voltage, HF current, HF power
Error display in safety field
Auto. limitation of the max. HF power
adjustable
Auto. limitation of the max. activation time
yes, display in safety field
Auto. function control
self-check after switching on the unit
Summary of Contents for 10128-002
Page 1: ...ERBOTOM ICC 200 Instruction manual 04 05 ERBE...
Page 2: ......
Page 8: ...1 2...
Page 10: ...2 2...
Page 18: ...3 8...
Page 19: ...International Standard...
Page 20: ......
Page 34: ...4 14...
Page 46: ...5 12 FORCED COAGULATION Power output as a function of the power limitation RL 350 ohms...
Page 56: ...7 2...
Page 62: ...9 2...
Page 64: ...10 2...
Page 66: ...11 2...
Page 72: ...12 6...