Basic setting and custom programming of the basic setting
If the unit is briefly (approx. 15 s) switched off and on again, the previous setting on the front
panel appears and the unit can be used again immediately. If the unit is switched off for longer
and switched on again, a freely programmable basic setting for the front panel appears, whereby
all displays 2.3, 3.6 blink and the unit cannot be activated until this basic setting is confirmed by
pressing any key. Then the displays are continuously illuminated and the unit can be used
immediately in this basic setting. If necessary, all settings on the front panel can now be changed.
After switching the unit off and on again after a longer period, the basic setting appears again. If
this basic setting proves to be inappropriate for the intended use of the unit, it can be individually
changed as follows.
1.
Switch off the unit.
2.
Press key 2.1 and switch on the power switch.
On display 2.3 appears
Pr
and on display 3.6 appears
1
, i.e. Test program 1 for changing
the basic setting has been called up.
3.
Press key 2.1. This activates Test program 1. The last stored basic setting appears. If however
no basic setting has been programmed by the user,
appears on displays 2.3, 3.6.
4.
Set the required basic setting.
5.
By switching off the unit, this front panel setting is automatically stored as the new basic
setting and appears automatically after every longer (approx. 15 s) period the unit is switched
off and on again.
Summary of Contents for 10128-002
Page 1: ...ERBOTOM ICC 200 Instruction manual 04 05 ERBE...
Page 2: ......
Page 8: ...1 2...
Page 10: ...2 2...
Page 18: ...3 8...
Page 19: ...International Standard...
Page 20: ......
Page 34: ...4 14...
Page 46: ...5 12 FORCED COAGULATION Power output as a function of the power limitation RL 350 ohms...
Page 56: ...7 2...
Page 62: ...9 2...
Page 64: ...10 2...
Page 66: ...11 2...
Page 72: ...12 6...