8-2
8.3
Automatic error documentation
Error numbers (ERROR No.) are assigned to the various errors recognized by the error recognition
system. If an error occurs, it is not only immediately reported visually and/or acoustically, but
also the corresponding error number is additionally stored in the unit, where it then also remains
stored when the unit is switched off. The last respective 10 error numbers can be called up at any
time as follows via Test program no. 2.
Calling up the test programs:
In the AUTO CUT field, press the EFFECT key with the unit switched off and
simultaneously switch on the power switch.
In the AUTO CUT display, Pr = Program appears,
In the AUTO COAG display, 1 = Test number 1 appears.
By pressing the
Ý
or
ß
key in the AUTO COAG field, the required test program number is
selected.
By pressing the EFFECT key in the AUTO CUT field, the required test program is started.
End the test program by briefly switching off the power switch.
Calling up the stored ERROR numbers
Call up Test program no. 2 (see above) and start.
By pressing the
Ý
or
ß
key in the AUTO COAG field, memory locations 1 to 10 are called
up.
In the AUTO CUT
display, the number of the memory location appears.
In the Auto COAG display, the error number appears.
The cause of error corresponding to each error number is listed in the error list (see above).
End this test program by briefly switching off the power switch.
Summary of Contents for 10128-002
Page 1: ...ERBOTOM ICC 200 Instruction manual 04 05 ERBE...
Page 2: ......
Page 8: ...1 2...
Page 10: ...2 2...
Page 18: ...3 8...
Page 19: ...International Standard...
Page 20: ......
Page 34: ...4 14...
Page 46: ...5 12 FORCED COAGULATION Power output as a function of the power limitation RL 350 ohms...
Page 56: ...7 2...
Page 62: ...9 2...
Page 64: ...10 2...
Page 66: ...11 2...
Page 72: ...12 6...