12
-3
Guidance and manufacturer's declaration - electromagnetic immunity
The equipment is intended for use in the electromagnetic environment specified below. The customer or the user of the
equipment should ensure that it is used in such an environment.
Immunity test
IEC 60601 test level
Compliance level
Electromagnetic
environment - guidance
Electrostatic discharge
(ESD) IEC 61000-4-2
±6 kV contact
±8 kV air
±6 kV contact
±8 kV air
Floors should be wood,
concrete or ceramic tile. If
floors are covered with
non-conductive synthetic
material, the relative hu-
midity should be at least
30%.
Electrical fast transient /
burst IEC 61000-4- 4
±2 kV for power supply
lines
±1 kV for input/output
lines
±2 kV for power supply
lines
±1 kV for input/output lines
Mains power quality
should be that of a typi-
cal commercial or hospi-
tal environment.
Surge IEC 61000-4-5
±1 kV differential mode
±2 kV common mode
±1 kV differential mode
±2 kV common mode
Mains power quality
should be that of a typi-
cal commercial or hospi-
tal environment.
Voltage dips, short interrup-
tions and voltage variations
on power supply input lines
IEC 61000-4-11
<5%
U
T
(>95% dip in
U
T
)
for 0.5 cycle
40%
U
T
(60% dip in
U
T
)
for 5 cycles
70%
U
T
(30% dip in
U
T
)
for 25 cycles
<5%
U
T
(>95% dip in
U
T
)
for 5 s
<5%
U
T
(>95% dip in
U
T
) for
0.5 cycle
40%
U
T
(60% dip in
U
T
) for 5
cycles
70%
U
T
(30% dip in
U
T
) for 25
cycles
<5%
U
T
(>95% dip in
U
T
) for
5 s
Mains power quality should
be that of a typical com-
mercial or hospital envi-
ronment. If the user of the
equipment requires contin-
ued operation during power
mains interruptions, it is
recommended that the
equipment be powered
from an uninterruptible
power supply or a battery.
Power frequency (50/ 60
Hz) magnetic field IEC
61000-4-8
3 A/m
3 A/m
Power frequency magnetic
fields should be at levels
characteristic of a typical
location in a typical com-
mercial or hospital envi-
ronment.
Note:
U
T
is the a.c. mains voltage prior to application of the test level.
Summary of Contents for 10128-002
Page 1: ...ERBOTOM ICC 200 Instruction manual 04 05 ERBE...
Page 2: ......
Page 8: ...1 2...
Page 10: ...2 2...
Page 18: ...3 8...
Page 19: ...International Standard...
Page 20: ......
Page 34: ...4 14...
Page 46: ...5 12 FORCED COAGULATION Power output as a function of the power limitation RL 350 ohms...
Page 56: ...7 2...
Page 62: ...9 2...
Page 64: ...10 2...
Page 66: ...11 2...
Page 72: ...12 6...