5-14
NESSY
Dependence of the warning signals on contact resistance R
ü
between the two contact surfaces of
a divided neutral electrode and the HF current I
HF
flowing through the entire surface of the
neutral electrode:
A = In this range, I
HF
is sufficiently small relative to R
ü
. No warning signal.
B = In this range, I
HF
is too large relative to R
ü
. The HF generator can be activated, however a
red warning signal and four warning tones are given after each activation of the HF generator.
In this condition, the unit must continue to be used only in case of emergency.
C = In this range, R
ü
is greater than 120 ohms. The HF generator cannot be activated. After
every activation attempt, a red warning signal and warning tones are given.