4-5
3
AUTO COAG function field
All parameters can be set in this function field that
are relevant to coagulation:
3.1
Selection of the coagulation mode
By pressing this key, one of the following coagulation modes can be selected:
3.2
Bipolar coagulation with or without AUTO START
3.3
Soft coagulation
3.4
Forced coagulation (Check the version. See Chapter 6.12)
3.5
Power limitation
The HF power output can be limited in BIPOLAR, SOFT and FORCED COAG mode in
1 watt steps from 120 watts to 1 watt. The HF power output can be limited in ARGON
COAG mode from 99 watts to 1 watt. If the display shows “—“, the Auto Coag mode is
switched off.
3.6
Display of the set power limitation in max. watt
3.7
AUTO START
ON / OFF for bipolar coagulation. The AUTO START function can only be switched on
with the setting AUTO COAG BIPOLAR.
WARNING!
Only use Auto Start if unintentional contact of tissue with the coagulation instrument can be
safely avoided. For endoscopic interventions, such as laparoscopy, pelviscopy or thoracoscopy,
Auto Start should not be used because unintentional contact of the coagulation instrument with
tissue cannot be safely avoided here.
In some countries the ICC units are supplied with the AUTO START function deactivated. This
may also apply in your case. If you nevertheless wish to use the AUTO START function, please
consult your local ERBE branch office. You will find the address on the last page of the Instruction
Manual. Technical Service will activate the AUTO START function for you on request.
The AUTO START key has no function on the ICC 200 No. 10128-027, 10128-028, 10128-058!
ARGON PLASMA COAGULATION
Set AUTO COAG to FORCED. Press the selection key again. The display shows: A.60. This
means: APC mode, power limitation 60 W. The APC mode is only available on the ICC 200 EA.
Summary of Contents for 10128-002
Page 1: ...ERBOTOM ICC 200 Instruction manual 04 05 ERBE...
Page 2: ......
Page 8: ...1 2...
Page 10: ...2 2...
Page 18: ...3 8...
Page 19: ...International Standard...
Page 20: ......
Page 34: ...4 14...
Page 46: ...5 12 FORCED COAGULATION Power output as a function of the power limitation RL 350 ohms...
Page 56: ...7 2...
Page 62: ...9 2...
Page 64: ...10 2...
Page 66: ...11 2...
Page 72: ...12 6...