Envent Engineering Ltd – 131S & 132S Software Manual – Revision 1
Page 24
4.6
Components
The components page contains parameters and configuration vital to the
calibration and operation of the gas chromatograph. Adjustment of the values on
this page will have an impact on the accuracy and reporting of the analyzer.
Proper care must be taken when making manual adjustments to the calibration
factors.
Top Half of Page
The top half of this page is where the calibration tables and component split
tables can be defined and configured. See below for a description of each
configurable field.
Sensor
Defines on which sensor the component will be measured
Component
Drop-down selecting the component, defined in
Component Properties
Cal. Gas Units
Selectable units which appear on the calibration standard
Cal. Gas
Defines the value which appears on the calibration
standard
Response Factor
Calibration factor calculated during calibration
Response Factor
Deviation
User defined deviation allowance for determining a
calibration PASS/FAIL